Description
The PXI-5122 is a high-performance digitizer featuring a digitizer speed of 100 MS/s. This device comes with several memory options to cater to various application needs, including 8 MB (778756-01), 32 MB (778756-02), 256 MB (778756-03), and 512 MB (778756-04). It utilizes 14-bit Analog-to-Digital Converters (ADCs) to ensure high-fidelity data conversion, achieving a Spurious Free Dynamic Range of 75 dBc and a Signal-to-Noise Ratio of 62 dB.
The input impedance is software selectable between 50 Ω or 1 MΩ, providing flexibility for various signal types. The device supports an input range from 200 mVpp to 20 Vpp, accommodating a wide range of signal amplitudes. It offers seven trigger modes, enabling users to precisely control the timing of their measurements. The front panel includes three BNC female connectors, a 9-pin mini-circular DIN, and two SMB jacks, offering versatile connectivity options.
Designed to operate within a temperature range of 0 to 55 °C, the PXI-5122 ensures reliable performance in various environments. It features self-calibration, with an external calibration recommended every two years to maintain accuracy. The device is compatible with NI-SCOPE driver, LabVIEW, and LabWindows/CVI software, ensuring easy integration into existing systems. A recommended warm-up time of 15 minutes is suggested for optimal performance.
For calibration, users have the option of NIST accredited calibrations through a third party or basic in-house calibration, providing flexibility in maintaining measurement accuracy. The maintenance phase for the PXI-5122 began on December 30, 2022
| Specification | Detail |
|---|---|
| Model | PXI-5122 |
| Digitizer Speed | 100 MS/s |
| Memory Options | 8 MB (778756-01), 32 MB (778756-02), 256 MB (778756-03), 512 MB (778756-04) |
| Analog-to-Digital Converters (ADCs) | 14-bit |
| Spurious Free Dynamic Range | 75 dBc |
| Signal-to-Noise Ratio | 62 dB |
| Input Impedance | Software selectable 50 Ω or 1 MΩ |
| Input Range | 200 mVpp to 20 Vpp |
| Trigger Modes | Seven |
| Front Panel Connectors | 3 BNC female, a 9-pin mini-circular DIN, and 2 SMB jack |
| Operating Temperature Range | 0 to 55 °C |
| Calibration | Self-calibrated; External calibration recommended every two years |
| Software Compatibility | NI-SCOPE driver, LabVIEW, LabWindows/CVI |
| Recommended Warm-Up Time | 15 minutes |
| Calibration Options | NIST accredited calibrations through a 3rd party; Basic in-house calibration |
| Maintenance Phase Start Date | December 30, 2022 |
Question 1: What is the digitizer speed of the PXI-5122?
Answer 1: 100 MS/s
Question 2: What is the ADC bit resolution of PXI-5122?
Answer 2: 14-bit
Question 3: What input impedance options does the PXI-5122 offer?
Answer 3: 50 Ω or 1 MΩ
Question 4: What is the SFDR rating for the PXI-5122?
Answer 4: 75 dBc
Question 5: How often is external calibration recommended for PXI-5122?
Answer 5: Every two years