img

199342B-01L

The PXI-5600 is a 3U PXI module RF downconverter with a 20 MHz bandwidth and a frequency range of 9 kHz to 2.7 GHz.

In Stock | Ships Today

img

    Description

    The Part Number 778283-01 specifies a RF Downconverter, specifically the model PXI-5600. This device is designed as a 3U PXI Module, tailored to fit seamlessly within the PXI ecosystem. It boasts a real-time bandwidth of 20 MHz, ensuring robust performance for a wide range of RF applications.

    Operating across a frequency range from 9 kHz to 2.7 GHz, the PXI-5600 is versatile and can be deployed in various scenarios requiring precise signal conversion. It features a typical noise density of < -140 dBm/Hz, contributing to its high sensitivity and efficiency in signal processing tasks. Additionally, its spurious-free dynamic range is stated to be > 80 dB, allowing for the accurate detection and measurement of signals amidst potential interference.

    The typical phase noise of the module is < -94 dBc/Hz at 10 kHz offset, indicating a low level of phase fluctuation and thus, high signal purity. With a timebase stability of ±20 ppb and a timebase accuracy of ±50 ppb, this module ensures precise timing and synchronization within your measurements.

    For software integration, the PXI-5600 requires NI-RFSA 2.1, NIPXI-5600 Support Driver, LabVIEW, LabWindows/CVI, and Microsoft C/C++, providing users with a range of options for application development and data analysis. The module is equipped with multiple connectors including FREQ REF IN, 10 MHz OUT, PXI 10 MHz I/O, OUTPUT, and INPUT, offering flexibility in setup and connectivity.

    The warm-up time for the PXI-5600 is set at 20 minutes, after which it operates at its optimal performance level. To maintain accuracy and reliability, a <

    Specification Detail
    Part Number 778283-01
    Category RF Downconverters
    Model PXI-5600
    Form Factor 3U PXI Module
    Real-Time Bandwidth 20 MHz
    Frequency Range 9 kHz to 2.7 GHz
    Typical Noise Density < -140 dBm/Hz
    Spurious-Free Dynamic Range > 80 dB
    Typical Phase Noise < -94 dBc/Hz at 10 kHz offset
    Timebase Stability ±20 ppb
    Timebase Accuracy ±50 ppb
    Software Requirements NI-RFSA 2.1, NIPXI-5600 Support Driver, LabVIEW, LabWindows/CVI, Microsoft C/C++
    Connectors FREQ REF IN, 10 MHz OUT, PXI 10 MHz I/O, OUTPUT, INPUT
    Warm-Up Time 20 minutes
    Calibration Period Annually or

    Question 1: What is the bandwidth of the PXI-5600?
    Answer 1: 20 MHz

    Question 2: What frequency range does the PXI-5600 cover?
    Answer 2: 9 kHz to 2.7 GHz

    Question 3: What is the typical noise density of the PXI-5600?
    Answer 3: < -140 dBm/Hz Question 4: How long is the warm-up time for the PXI-5600?
    Answer 4: 20 minutes

    Question 5: What software is required for the PXI-5600?
    Answer 5: NI-RFSA 2.1, Driver, LabVIEW, LabWindows/CVI, C/C++