Description
The product with the part number 777060-01 falls under the category of PXI Multifunction DAQ and is designated as model PXI-6070E. This model is capable of achieving a maximum sampling rate of 1.25 MS/s and boasts a resolution of 12-bit. It offers a choice between 16 or 64 single-ended analog inputs, allowing for adaptable configuration based on the specific needs of the project or experiment at hand.
In terms of output capabilities, the model provides two 12-bit analog outputs and includes two 24-bit, 20 MHz counter/timers. Additionally, it is equipped with eight digital I/O lines and utilizes a 68-pin male 0.050 D-type I/O connector. The device offers functional isolation up to 250 V, ensuring safety and reliability in various operating environments.
Compatibility with software is crucial for effective data management and analysis. The PXI-6070E is compatible with several software platforms including NIDAQmx, LabVIEW, LabWindows/CVI, and Measurement Studio. Before initial use, it requires a warm-up time of 15 minutes, and to maintain its precision, calibration is recommended at least once a year.
The physical characteristics of the device include a weight of 7.1 oz and dimensions of 6.3 x 3.9 in. It is designed to operate within a temperature range of 0 to 55 °C and can be stored safely at temperatures ranging from –20 to 70 °C. The device maintains its integrity under relative humidity levels of 10 to 90% (non-condensing). Lastly, it conforms to IEC 61010 safety standards, ensuring that it meets international requirements for electrical safety.
| Specification | Detail |
|---|---|
| Part Number | 777060-01 |
| Category | PXI Multifunction DAQ |
| Model | PXI-6070E |
| Max Sampling Rate | 1.25 MS/s |
| Resolution | 12-bit |
| Analog Inputs | 16 or 64 single-ended |
| Streaming to Disk Rate | Up to 1.25 MS/s |
| Analog Outputs | Two 12-bit |
| Counter/Timers | Two 24-bit, 20 MHz |
| Digital I/O Lines | Eight |
| I/O Connector | 68-pin male 0.050 D-type |
| Functional Isolation | 250 V |
| Software Compatibility | NIDAQmx, LabVIEW, LabWindows/CVI, Measurement Studio |
| Warm-up Time | 15 minutes |
| Calibration Frequency | At least once a year |
| Weight | 7.1 oz |
| Dimensions | 6.3 x 3.9 in. |
| Operating Temperature | 0 to 55 °C |
| Storage Temperature | –20 to 70 °C |
| Relative Humidity | 10 to 90% (non-condensing) |
| Safety Standards | IEC 61010
Question 1: What is the max sampling rate of the PXI-6070E? Question 2: How many analog inputs does the PXI-6070E support? Question 3: What software is the PXI-6070E compatible with? Question 4: What is the recommended calibration frequency for the PXI-6070E? Question 5: What safety standard does the PXI-6070E conform to? |