img

200416A-01

This device, with part numbers 783130-01, 154888C-02L, features a 7½-digit capacity, voltage range of ±100 mV to ±1000 V, and can sample at rates from 10 S/s to 1.8 MS/s.

In Stock | Ships Today

img

    Description

    The product, with part numbers 783130-01, 154888C-02L, features a precise 7½-Digit digit capacity, allowing for highly accurate measurements. It supports a wide voltage range from ±100 mV to ±1000 V (DC or AC coupled), making it versatile for various electrical measurement needs. Additionally, the device can measure current within the range of ±1 µA to ±3 A, offering flexibility in current sensing applications.

    The sample rate range of this product spans from 10 Samples per second (S/s) to 1.8 Mega Samples per second (MS/s), catering to both slow and fast signal measurements efficiently. It is designed with an input impedance of 10 MΩ ± 2% in parallel with 90 pF, ensuring accurate signal capture without significant loading effects. For applications involving DC voltage, it can tolerate up to a maximum DC voltage component of 400 V.

    This device is capable of operating over a frequency range of 15 Hz to 500 KHz, suitable for a broad spectrum of frequency analysis. It also features a period measurement range from 2 µs to 66.67 ms, allowing for precise timing measurements. The ADC linearity is exceptional, with up to 0.5 ppm of reading + 0.5 ppm, ensuring high fidelity in digital conversion.

    For optimal performance, it requires a warm-up time of 60 minutes and should undergo calibration once every 2 years to maintain measurement accuracy. The input protection is rated at 1000 VDC or Vpk, providing safety against high voltage spikes. Lastly, the current mode fuse rating is specified as T 3.5 A 1000 V, time-lag Minimum interrupt rating: 10 kA Siba, offering robust protection against overcurrent conditions.

    Specification Description
    Part Number 783130-01, 154888C-02L
    Digit Capacity 7½-Digit
    Voltage Range ±100 mV to ±1000 V (DC or AC coupled)
    Current Range ±1 µA to ±3 A
    Sample Rate Range 10 S/s to 1.8 MS/s
    Input Impedance 10 MΩ ± 2% in parallel with 90 pF
    Maximum DC Voltage Component 400 V
    Frequency Range 15 Hz to 500 KHz
    Period Measurement Range 2 µs to 66.67 ms
    ADC Linearity Up to 0.5 ppm of reading + 0.5 ppm
    Warm-up Time 60 minutes
    Calibration Period Once every 2 years
    Input Protection 1000 VDC or Vpk
    Current Mode Fuse Rating T 3.5 A 1000 V, time-lag Minimum interrupt rating: 10 kA Siba

    Question 1: What is the digit capacity of the PXIe-4081?
    Answer 1: 7½ digits

    Question 2: What voltage range does the PXIe-4081 support?
    Answer 2: ±100 mV to ±1000 V

    Question 3: What is the sample rate range of the PXIe-4081?
    Answer 3: 10 S/s to 1.8 MS/s

    Question 4: How often does the PXIe-4081 need calibration?
    Answer 4: Every 2 years

    Question 5: What is the maximum DC voltage component for PXIe-4081?
    Answer 5: 400 V