The PXI Programmable Power Instruments include models such as the PXIe-4112, which is identified by part numbers 782857-01 and 782956-01. This model boasts 2 channels, making it versatile for a range of applications.
The voltage and current capabilities of the PXIe-4112 are quite notable. It offers a voltage range up to 60 V and a current range up to 1 A. Moreover, the minimum programmable voltage level/limit is as low as 0.1 V, and for current, it goes down to 0.01 A. These features ensure precise control over power delivery to the target system or device.
When it comes to output power, each channel of the PXIe-4112 can deliver up to 60 W, totaling 120 W for the unit. This is complemented by an analog output voltage range that stretches from 0 V to 60 V, providing a wide spectrum for various test scenarios.
The device’s auxiliary power input line regulation is another point of excellence, with voltage regulation at 1 mV and current regulation at 0.2 mA. Additionally, the remote sense feature can accommodate a maximum output lead drop of 1 V per lead and a maximum sense lead resistance of 1 Ω per lead, ensuring the accuracy of power delivery even with some resistance in the leads.
For power requirements, the PXIe-4112 draws 0.5 A from a 3.3 V rail and 0.25 A from a 12 V rail. The auxiliary power source input requirements are also specific, needing 45.6 VDC to 50.4 VDC with a maximum current of 3.5 A. These specifications highlight the device’s efficient power usage and compatibility with standard power supplies.
| Specification | Details |
|---|---|
| Part Numbers | 782857-01, 782956-01 |
| Category | PXI Programmable Power Instruments |
| Model | PXIe-4112 |
| Channels | 2 |
| Voltage Range | 60 V |
| Current Range | 1 A |
| Minimum Programmable Voltage Level/Limit | 0.1 V |
| Minimum Programmable Current Level | 0.01 A |
| Output Power | 60 W/channel, Total 120 W |
| Analog Output Voltage Range | 0 V to 60 V |
| Auxiliary Power Input Line Regulation | Voltage: 1 mV, Current: 0.2 mA |
| Remote Sense | Max output lead drop: 1 V per lead, Max sense lead resistance: 1 Ω per lead |
| Power Requirements | 0.5 A from 3.3 V rail, 0.25 A from 12 V rail |
| Auxiliary Power Source Input Requirements | 45.6 VDC to 50.4 VDC, 3.5 A max |