Description
The DAQCard-5102 is a PC-based oscilloscope/digitizer part number 183442C-01, manufactured by National Instruments. This device is designed for use with a PCMCIA bus interface, making it ideal for portable applications where a compact and efficient solution is required. It features 2 input channels with an 8-bit resolution, capable of accommodating a variety of signals.
For triggering, the DAQCard-5102 includes 1 analog and 2 digital triggers, providing flexibility in signal capture and analysis. The input bandwidth is rated at 15 MHz, suitable for a wide range of applications from portable field test to high-speed waveform capture and electronic testing. The device operates with a sampling rate range from 20 MS/s to 1 kS/s, and it has a memory depth of 662 ksamples, allowing for extensive data collection and detailed waveform analysis.
The oscilloscope/digitizer offers software-selectable AC/DC coupling, and its analog channel trigger features include hysteresis, as well as software-selectable level and slope. Digital triggers are of the standard TTL/CMOS type, providing compatibility with a wide range of digital devices.
Regarding software and operating system compatibility, the DAQCard-5102 requires the NI-Scope driver and is compatible with Measurement Studio, LabWindows, LabVIEW, Visual Basic, and C/C++. It supports Windows NT/2000/XP operating systems, ensuring easy integration into existing setups.
This device is particularly suitable for applications such as portable field test, power supply testing, high-speed waveform capture, transmission line testing, and general electronic testing. It has a warm-up time of 15 minutes and a typical power consumption of 5 VDC, 260 mA, making it an efficient choice for both field and lab work
| Specification | Details |
|---|---|
| Part Number | 183442C-01 |
| Category | PC-based Oscilloscopes/Digitizers |
| Model | DAQCard-5102 |
| Manufacturer | National Instruments |
| Bus Interface | PCMCIA |
| Input Channels | 2 |
| Resolution | 8 bits |
| Triggers | 1 analog and 2 digital |
| Input Bandwidth | 15 MHz |
| Required Driver | NI-Scope |
| Software Compatibility | Measurement Studio, LabWindows, LabVIEW, Visual Basic, C/C++ |
| Operating Systems Compatibility | Windows NT/2000/XP |
| Sampling Rate | 20 MS/s to 1 kS/s |
| Memory Depth | 662 ksamples |
| Coupling | Software-selectable AC/DC |
| Analog Channel Trigger Features | Hysteresis, software-selectable level and slope |
| Digital Trigger Types | Standard TTL/CMOS |
| Applications | Portable field test, power supply testing, high-speed waveform capture, transmission line testing, electronic testing, etc. |
| Warm-up Time | 15 minutes |
| Power Consumption | 5 VDC, 260 mA typical |
Question 1: What is the interface type of DAQCard-5102?
Answer 1: PCMCIA
Question 2: How many input channels does the DAQCard-5102 have?
Answer 2: 2 input channels at 8 bits resolution
Question 3: What is the input bandwidth of the DAQCard-5102?
Answer 3: 15 MHz
Question 4: What is the maximum sampling rate of DAQCard-5102?
Answer 4: 20 MS/s
Question 5: Is the DAQCard-5102 compatible with Windows XP?
Answer 5: Yes