img

198854E-03L

The NI-9213 is a C Series Analog Input model with 16 thermocouple channels, ±2.25°C accuracy, and supports various programming environments.

In Stock | Ships Today

img

    Description

    The NI-9213 is a C Series Analog Input device designed specifically for high accuracy temperature measurements. This model is identified by several Part Numbers including 780493-01, 780493-02, 785185-01, 785185-02, and 198854E-03L, catering to a wide range of applications and compatibility requirements. The NI-9213 stands out with its Conformal Coated, 16 Channel, Thermocouple Input, providing robustness and reliability for critical measurements.

    It has 16 channels, allowing for extensive data collection and monitoring across multiple points. The device boasts a Maximum Accuracy of ±2.25°C, ensuring precise readings critical for quality control and research purposes. With a Sample Rate of 75 Samples per second, it offers a balanced performance between speed and resolution.

    The NI-9213 is designed with safety and durability in mind, featuring 250 Vrms Ch-Earth continuous isolation to protect both users and equipment. Connection is made simple and secure through spring terminals, facilitating easy setup and maintenance. Additionally, it includes filtering capabilities for 50 Hz and 60 Hz noise, enhancing signal quality in environments prone to electrical interference.

    At a weight of just 5.6 oz, it is a lightweight yet powerful device. It supports both high-resolution and high-speed timing modes, making it versatile for different types of experiments and data acquisition tasks. The NI-9213 is compatible with a range of software including LabVIEW, LabWindows/CVI, Microsoft Visual C++, Microsoft Visual Basic, Borland C++, and Microsoft .NET, ensuring seamless integration into existing systems.

    The recommended Calibration Interval is at least once a year to maintain accuracy and reliability. It operates within a wide Operating Temperature range of -40°C to 70°C, making it suitable for use in extreme conditions.

    Question 1: What is the max accuracy of the NI-9213?
    Answer 1: ±2.25°C

    Question 2: How many thermocouple channels does the NI-9213 have?
    Answer 2: 16 channels

    Question 3: What is the sample rate of the NI-9213?
    Answer 3: 75 Samples per second

    Question 4: What is the weight of the NI-9213?
    Answer 4: 5.6 oz

    Question 5: What isolation level does the NI-9213 feature?
    Answer 5: 250 Vrms Ch-Earth continuous isolation

    Specification Detail
    Part Numbers 780493-01, 780493-02, 785185-01, 785185-02, 198854E-03L
    Category C Series Analog Input
    Model NI-9213
    Description Conformal Coated, 16 Channel, Thermocouple Input
    Channels 16
    Maximum Accuracy ±2.25°C
    Sample Rate 75 Samples per second
    Isolation 250 Vrms Ch-Earth continuous
    Terminals Spring terminal
    Filtering 50 Hz and 60 Hz noise
    Weight 5.6 oz
    Timing Modes High-resolution and high-speed
    Software Compatibility LabVIEW, LabWindows/CVI, Microsoft Visual C++, Microsoft Visual Basic, Borland C++, Microsoft .NET
    Calibration Interval Recommended at least one year
    Operating Temperature -40°C to 70°C