Agilent 60501B Electronic Load Module
The 60501B Electronic Load Module has a current between 0 to 60 A, Voltage between 3V to 60 V and a power of 300 W at 40º. This model has a constant current mode range between 0 to 6 A, a constant resistance mode range between 0.033 to 1 Ω and a constante voltage mode range between 0 to 60 V.
Agilent dc electronic loads are perfect for the test and assessment of DC control sources and power segments and are appropriate for applications in many areas, such as assembling, innovative work, furthermore, approaching investigation.
These dc electronic loads respond to directions from the industry-standard SCPI order set. Besides, the components of these dc electronic loads are completely compatible with one another. For instance, test programs produced for 6060B 300 W single input electronic load or 60502B 300 W single-input load module are interchangeable.
Since Agilent electronic loads highlight an incorporated GPIB programmer, pulse generator, current shunt, DMM, and cabling, their execution is indicated as a system. The specifications cover all the coordinated capacities as one unit, which takes out the need to figure the genuine execution of the robotized test framework in light of every segment's determination. The one-box arrangement of this model makes the mix and documentation of your test system quick and simple. This model also includes six unstable client quantifiable states that enable you to effortlessly spare settings for later review and an additional user-definable power-up state that allows you to characterize settings that are recollected when the unit is switched off and then recalled when it is switched on again
The Agilent 60501B module alongside the dc electronic load centralized computer are perfect for the test and assessment of dc control sources and power segments and are appropriate for applications in many areas, such as innovative work, production and approaching examination. Every module incorporates programming inputs that permit control of load current through a simple control voltage. This model also includes a simple voltage control in steady current mode, full security from over-voltage, over-current, over-temperature and reverse polarity.
This model can work in three modes: constant current mode, constant voltage mode (ranges between 0.067 Ω to 2 Ω ) and constant resistance mode. The 60501B has a current between 0 to 30 A, 3 to 60 V (min dc operation from 0 to 2 V for 0 to 30 A) and 150 Watts at 40ºC derated to 112 Watts at 55ºC of power.
Agilent single-input loads and load centralized servers are outfitted with standard GPIB interfaces. This implicit IEEE-488 interface permits finish control of all load capacities and in addition readback of info for voltage, current, control, and itemized working status. Each stand alone load or load module likewise incorporates programming inputs that permit control of load current via an analog voltage. Other framework highlights adding to the one-box arrangement idea are interior voltage and current screens and an interior transient generator with programmable amplitudes, recurrence, duty cycle and slew. The one-box arrangement spares space, cost, and time while making these dc electronic loads simple to incorporate into mechanized test frameworks.
60501B is a electronic load module. It is a part of 6050A or 6051A Multiple-Input electronic Load Mainframes. Its capacity is 150 watts.
Each module can operate in three modes.
1. Constant current (CC) mode
2. Constant voltage mode (CV)
3. Constant resistance mode (CR).
Technical Specifications
Specifications:
Current |
0 to 30 A |
Voltage |
3 to 60 V (min dc operation from 0 to 2 Volts for 0 to 30 A) |
Power |
150 Watts at 40 °C derated to 112 Watts at 55 °C |
Constant Current Mode:
Ranges |
0 to 3 A, and 0 to 30 A |
Accuracy |
(After 30 seconds delay): ±0.1% ±40 mA (both ranges) |
Resolution |
0.8 mA (3 A range); 8 mA (30 A range); |
Regulation |
10 mA (both ranges) |
Temperature Coefficient |
100 ppm/ °C ±3 mA/ °C (both ranges) |
Constant Resistance Mode:
Ranges |
0.067 to 2 Ω; 2 Ω to 2 kΩ; 2 kΩ to 10 kΩ; |
Accuracy |
±0.8% ±16 mΩ with ≥ 6 A at input (2 Ω range); ±0.3% ±5 mS with ≥ 6 V at input (2 kΩ & 10 kΩ ranges); |
Resolution |
0.54 mΩ (2 Ω range); 0.14 mS (2 kΩ range); 0.014 mS (10 kΩ range); |
Regulation |
10 mV with remote sensing (2 Ω range); 10 mA (2 kΩ & 10 kΩ ranges); |
Temperature Coefficient |
800 ppm/ °C ±0.8 mΩ/ °C (2 Ω range); 300 ppm/ °C ±0.5 mS/ °C (2 kΩ & 10 kΩ ranges); |
Constant Voltage Mode:
Ranges |
0 to 60 V |
Accuracy |
±0.1% ±50 mV |
Resolution |
16 mV |
Regulation |
5 mV (remote sensing); 40 mV (Local sensing) |
Temperature Coefficient |
100 ppm/ °C ±5 mV/ °C |