The NI USB-6259 is a 16-bit PC-based multifunction DAQ with a maximum sample rate of 1.25 MS/s, supporting analog and digital triggering.

In Stock | Ships Today

img

    Description

    The product with the part number 779628-12, which includes specific versions such as 780114-01, 780114-04, and 779628-01, falls under the category of PC-based Multifunction DAQ. The model of this product is USB-6259. It is described as an NI USB-6259 M Series DAQ Device, featuring a screw terminal and is supported by the NI-DAQMX Driver, specifically designed for use in Taiwan with a 110V power supply. This device boasts a resolution of 16-bit, ensuring high-quality data acquisition.

    It supports various triggering types including Analog Triggering and Digital Triggering, providing flexibility in data capture methods. The device is equipped with two 32-bit, 80 MHz counters/timers, enhancing its capability in time-sensitive applications. With regard to input and output channels, it offers 32 single-ended or 16 differential analog input channels and 4 analog output channels. Furthermore, it is capable of a maximum sample rate of 1.25 MS/s per channel, which allows for the capture of fast-changing signals.

    The USB-6259 supports four high-speed, bidirectional data streams, making it suitable for applications requiring high data throughput. The accuracy of measurements is ensured through NI-MCal technology, which provides advanced calibration for the device. It is compatible with several operating systems including Windows Vista, Windows XP, and Windows 2000, ensuring flexibility in terms of system integration. The device requires a warm-up time of 30 minutes and should be calibrated at least once every 2 years to maintain accuracy.

    It offers a comprehensive range of analog and digital I/O including 32 analog input channels, 4 analog output channels, and multiple clocked digital lines, catering to a wide range of data acquisition needs. The signal conditioning compatibility with NI-SCC allows for enhanced signal processing capabilities. Connectivity is facilitated through a USB 2.0 Hi-Speed or Full-Speed interface, ensuring easy integration with various computer systems. Lastly, the physical dimensions of the screw terminal version are 10.5 x 6.73 x 1.75 inches, making it a compact and versatile solution for data acquisition tasks.

    Specification Detail
    Part Number 779628-12 (with specific versions: 780114-01, 780114-04, 779628-01)
    Category PC-based Multifunction DAQ
    Model USB-6259
    Description NI USB-6259 M Series DAQ Device, Screw Term, NI-DAQMX Driver, Taiwan (110V)
    Resolution 16-bit
    Triggering Types Analog Triggering, Digital Triggering
    Counters/Timers Two 32-bit, 80 MHz
    Analog Input Channels 32 single-ended or 16 differential
    Maximum Sample Rate (per channel) 1.25 MS/s
    Analog Output Channels 4
    Data Streams Four High-Speed, Bidirectional
    Accuracy Technology NI-MCal
    Operating Systems Windows Vista, Windows XP, Windows 2000
    Warm-up Time 30 minutes
    Calibration Interval At least once every 2 years
    Analog and Digital I/O 32 analog input, 4 analog output, multiple clocked digital lines
    Signal Conditioning Compatibility NI-SCC
    USB Interface USB 2.0 Hi-Speed or Full-Speed
    Physical Dimensions (Screw Terminal) 10.5 x 6.73 x 1.75 inches

    Question 1: What is the sample rate of the NI USB-6259?
    Answer 1: 1.25 MS/s per channel

    Question 2: How many analog input channels does the USB-6259 have?
    Answer 2: 32 single-ended or 16 differential

    Question 3: What operating systems is the USB-6259 compatible with?
    Answer 3: Windows Vista, XP, 2000

    Question 4: What is the required calibration interval for the USB-6259?
    Answer 4: At least once every 2 years

    Question 5: Does the USB-6259 support digital triggering?
    Answer 5: Yes