Description
“`html
The Part Number 783129-01 refers to the PXIe-4080 model, a high-performance 6½ digit DMM (Digital Multimeter) and 300 V Digitizer designed for advanced measurement capabilities. This model is engineered to deliver precise voltage measurements within a range of ±300 V and supports a digitizer sample rate of up to 1.8 MS/s, allowing for high-speed data acquisition.
The PXIe-4080 boasts an input impedance of 1 MΩ ± 2% in parallel with 150 pF, ensuring accurate measurements across various applications. It can handle a maximum DC voltage component of 250 V and offers an impressive CMRR (Common Mode Rejection Ratio) of >70 dB from DC to 60 Hz, highlighting its ability to reject unwanted signals and noise.
Featuring a frequency range from 15 Hz to 500 KHz and a period measurement range from 2 µs to 66.67 ms, this model is versatile enough to accommodate a wide range of measurement tasks. The input protection is rated at 300 VDC or ACrms, safeguarding the device against potential damage from overvoltage.
The PXIe-4080 provides a resolution that is 10 to 23-bit flexible, allowing for adjustments based on the precision requirements of the measurement. It is capable of current measurements up to 1 A and resistance measurements up to 100 MΩ. Additionally, the current mode fuse rating is specified as T 1 A 400 V, with a minimum interrupt rating of 500 A (Littelfuse 0477001.MXP), ensuring the device’s safety and reliability.
With a <
Specification | Detail |
---|---|
Part Number | 783129-01 |
Model | PXIe-4080 |
Description | High-Performance 6 1/2 Digit DMM and 300 V Digitizer |
Digits | 6½ |
Voltage Measurement Range | ±300 V |
Digitizer Sample Rate | Up to 1.8 MS/s |
Input Impedance | 1 MΩ ± 2% in parallel with 150 pF |
Maximum DC Voltage Component | 250 V |
CMRR | >70 dB (DC to 60 Hz) |
Frequency Range | 15 Hz to 500 KHz |
Period Measurement Range | 2 µs to 66.67 ms |
Input Protection | 300 VDC or ACrms |
Resolution | 10 to 23-bit flexible |
Current Measurements | Up to 1 A |
Resistance Measurements | Up to 100 MΩ |
Current Mode Fuse Rating | T 1 A 400 V, time-lag user-replaceable Minimum interrupt rating: 500 A Littelfuse 0477001.MXP |
Power Consumption | Less than 9 W from PXI Express backplane |
Software Compatibility | NI-DMM driver, LabVIEW, LabWindows/CVI, ANSI C, Visual Basic |
Question 1: What is the max DC voltage component for PXIe-4080?
Answer 1: 250 V
Question 2: What is the digitizer sample rate of PXIe-4080?
Answer 2: Up to 1.8 MS/s
Question 3: What is the frequency range of the PXIe-4080?
Answer 3: 15 Hz to 500 KHz
Question 4: What is the current mode fuse rating of PXIe-4080?
Answer 4: T 1 A 400 V
Question 5: What is the CMRR of PXIe-4080 from DC to 60 Hz?
Answer 5: >70 dB